1. Overview
- Enables filtration, pre-heating, dearation, and precise dosing of media in the vessel within the dissolution bath
- 21 litre media tank capacity allows filling of dissolution vessels for upto 3 dissolution tests (n=6)
- Batch degassing ensures inter-vessel degassing consistency
- Volumetric dispensing with an accuracy better than 1% as per USP, FDA and Ph. Eur. requirement
- Special nozzle design allows the handling of surfactant and frothing media like Sodium Lauryl Sulphate (SLS)
- Programmable deaeration time, temperature and vacuum for effective deaeration of different media
- Maintains media under vacuum at desired temperature while in idle condition
- Virtually zero dead volume minimizes wastage of media
- Programmable dispensing volumes from 200 ml to 4 ltrs
- Fast filling of dissolution vessels at a dispensing rate of 900 ml/ min
- Programmable rinse and cleaning cycle allows the instrument to run through a prevalidated self cleaning procedure
- On- board memory stores 10 batch information of degassed media
- Unparalleled versatility due to 1 ml volume change increments
- 360 turning radius allows it to be maneuvered in tight places
- Multifunction single hose for dispensing, aspirating, cleaning, rinsing and draining ensures no lifting of heavy media tanks
- Ergonomically designed nozzle with embedded dispense switch for ease of operation
- User friendly validation and calibration of media volumes
- Print out of validation, calibration, dispensed volume and temperature reports
2. Features:
- Media Volume: Programmable from 0.2 litre to 21 litres
- Accuracy: ±1%
- Flow rate: 900 ml/min. (During run cycle)
- Dispense:
- Dispense Volume: Programmable from 0.2 L to 4 L
- Accuracy: Better than ±1%
- Flow rate: 900 ml/min.
- Temperature: Temperature Range: 20 o C to 50 o C
- Accuracy: ±0.5 o C
- Media preparation time: 45 min. (For 21 L of media @ambient temperature 25 o C)
- O 2 level: Normal tap water which contains 85% of dissolved O 2 before degassing will contain 40% of dissolved O 2 after degassing (Measured by DO 2 meter)
- Hose: Single (For aspiration, dispense and drain)
- Media tank: S.S. 316, Teflon coated
- Vacuum pump: 760 mmHg
- Piston pump: Ceramic metering piston pump
- Accuracy: Better than ±1%
- Power: 1200 watts
- Power supply: 230V AC/50 Hz. ±10%, : 100/110 VAC, 50/60 Hz, 550VA (on request)
- Ambient conditions:
- Humidity: 70% max.
- Storage temperature: 10 o C to 50 o C
- Operating temperature: 10 o C to 30 o C
- Dimensions: (ǿ 554 x 920 H) mm
- Weight: Net 100 kgs